The room-use section of a report: the operating room's electricity during each procedure, on the national grid of the hospital's country.
Methods of ClinImpact 2.21. Reports name the version they were computed with; the changelog says what changed.
Room use is Σ room minutes ÷ 60 × the room type's kWh per hour × the low-voltage grid factor of the hospital's country.
A procedure's room minutes are its room time (from the patient entering the operating room to leaving it) when that is recorded, and its procedure time otherwise. A procedure without a room or a time is named on a "not counted" line.
A room can instead carry a week's energy and the number of procedures in that same week. The kWh per procedure is then the one divided by the other, which counts waiting and turnover time, and it replaces kWh per hour × duration for that room.
Equipment#
Since 2.21.0 a project can name the devices its procedures use (an ultrasound, an endoscopy tower, bipolar diathermy, a tourniquet) with their kWh per procedure, for example the power on the datasheet times the time it runs. Each is a line "Equipment > device": kWh per procedure × the procedures recorded × the same low-voltage grid factor, under the Room part and split evenly over the procedures. The room's own kWh may already include its equipment; the same energy must not be entered twice.