The sterilisation section of a report: the tap water, RO water and electricity of the washers and the autoclave, divided by the DIN baskets each cycle carries and multiplied by the baskets each procedure sent.
Methods of ClinImpact 2.21. Reports name the version they were computed with; the changelog says what changed.
Each hospital records, for the washer, the cart washer and the autoclave, the tap water, RO water and electricity of one cycle or charge, and the number of instrument sets (DIN) it carries.
Each procedure records the DIN baskets it sent through sterilisation, which may differ by surgeon, patient or case. The project's own figure is the default a new procedure starts from, and it is counted for a procedure without a figure of its own. A project without procedures (a standard case, a component study) counts the project's figure once.
| Line | Formula | Dataset |
|---|---|---|
| Tap water | Σ machines (L per cycle ÷ DIN per cycle) × DIN used, 1 L = 1 kg | Tap water, at user {CH} |
| RO water | the same with RO litres | Water, deionised, water balance according to MoeK 2013, at plant {CH} |
| Electricity | Σ machines (kWh per cycle ÷ DIN per cycle) × DIN used | Electricity, low voltage, at grid {hospital country} |
DIN used is the sum over the procedures of each one's DIN baskets (the project's figure where a procedure has none); it is not the project's figure times the number of procedures.
These lines exist only when the hospital has sterilisation switched on. "No sterilisation needed" or "no reusable set used" sets DIN to 0.
RO water is costed with the deionised-water dataset. Costing it as tap water times the RO input multiplier would count the feed water twice. The multiplier, the STU capacity and the cost per DIN take no part in the figure. The settings are frozen with each published report and printed as the "Sterilisation calculation".
Not modelled:
- detergent and chemicals;
- steam and compressed air as separate inputs;
- wrap and containers;
- instrument wear and the set's manufacture (unless its instruments are listed among the products as reusable, see Methods: Products);
- transport to the sterilisation department.
Standby and test cycles#
Standby and test cycles are optional per hospital. The machines stay on between cycles and run a test cycle each week. Per machine type and per week:
- standby = standby power × (machines × 168 h − cycles per week × cycle time);
- test = machines × test cycles per week × kWh per test cycle.
The cycles' own energy is already in the kWh per cycle. The sum of standby and test, divided by the DIN of the week (cycles per working day × working days × DIN per cycle), gives an electricity line per DIN at the hospital's grid. If a figure is missing, nothing is added.
Detergent and products#
Since 2.21.0 the washer's detergent per cycle (in ml; for Aniosyme, the dose of 11 ml per 38 L times the litres of a cycle) is spread over the DIN of a cycle like its water, counting 1 ml as 1 g. BAFU:2026 has no dataset for a cleaning or enzymatic detergent, so "Soap, at plant {RER}" stands in as a proxy until one is chosen, and the line says so.
Products used by the sterilisation itself, such as the wraps around the nets, are template lines of their own part: counted per procedure like any product, and reported under the sterilisation rather than the operation.